When the standard is not the standard
Hidden concentration errors despite satisfactory chromatographic performance

A calibration standard can satisfy every chromatographic check the instrument monitors — and still be an order of magnitude from the concentration written on the flask.

Material
Coenzyme Q10 (ubidecarenone)
Diluent
Methanol:ethanol 65:35 (v/v)
Instrument
Reversed-phase HPLC, photodiode-array detection
First failure
Levels from a supersaturated stock, ~10-fold low
Second failure
Prepared below saturation, 82.3 % of nominal
System suitability
Passed throughout both
Licence
CC-BY 4.0 · Working Paper
Status
Preprint · not yet peer reviewed
01

What the paper reports

Two separate failures in one method development exercise, arriving at the same place by different routes, with the solubility measurements that explain them.

  • A stock prepared at 1 mg/mL — about five times the equilibrium solubility of coenzyme Q10 in that diluent — was forced into solution by sonication and warming to 36–37 °C, then held overnight. Levels made from it the next morning were roughly ten-fold low.
  • Through that entire sequence the instrument reported plate counts of 2,489–2,641, tailing factors of 1.272–1.296, injection repeatability of 0.23 % and carryover of 0.000 %. No chromatographic parameter indicated a fault.
  • Apparent equilibrium solubility was about 206 µg/mL at 24.5 °C and 119 µg/mL at refrigerator temperature — a fall of 42 %, which is why the conventional advice to refrigerate standards makes this failure more likely rather than less.
  • Dissolution from the solid took about 20 hours at 88 µg/mL and about 40 hours at 150 µg/mL, while crystallisation on cooling happened overnight. The process is slow in one direction and fast in the other.
  • The second failure was prepared below the solubility limit and judged visually clear after 20 hours. It assayed at 82.3 % of nominal. Attributed entirely to undissolved material, that is about 0.78 mg of the 4.4 mg weighed — an amount nobody would see in a flask.
  • Two minor peaks sharing the analyte chromophore, at relative retentions of 0.745 and 1.373, are consistent with coenzyme Q9 and coenzyme Q11, and were enriched in the saturated solution in the direction solubility predicts.
02

What we are not claiming

  • The deficit in the second preparation is established by the data. Its mechanism is not. Incomplete dissolution is the most likely explanation, not a demonstrated one.
  • Each solubility figure is a single determination, and both lie above the highest calibration level, so both were obtained by extrapolation beyond the calibrated range.
  • No certified reference material was used. Absolute concentrations rest on the gravimetric preparation rather than on an independent standard.
  • One analyst, one instrument, no inter-laboratory comparison. The identification of the minor peaks rests on retention behaviour and spectra, not on authentic standards.

The paper

When the standard is not the standard: hidden concentration errors in coenzyme Q10 calibration standards despite satisfactory chromatographic performance

Author
Kalpesh Upadhyay, Raupyam Labs Limited
ORCID
0009-0002-1619-9002
Posted
2 November 2026 · ChemRxiv
DOI
10.26434/chemrxiv.15011557/v1
Licence
CC-BY 4.0

The full text is on ChemRxiv, not reproduced here. It carries the instrument conditions, both calibration datasets, the solubility and dissolution measurements, the filtration control, the minor-peak assignment and the stated limitations, with supplementary peak tables.

Revisions

  1. 2026-11-02Preprint posted to ChemRxiv as a Working Paper under CC-BY 4.0, with supplementary data. DOI issued. Manuscript being prepared for journal submission.
  2. 2026-11-03Submitted to the Journal of AOAC INTERNATIONAL as a Research Article, manuscript AOAC-26-0277. Under editorial assessment. Not yet peer reviewed, and submission is not acceptance.

This page stays open. Review outcomes and corrections are added here rather than quietly edited in.

Raupyam Labs Limited. This page describes standard preparation and makes no claims regarding any finished product or any named supplier. Every figure in the paper was plotted from the instrument’s own exported data. This work has not been peer reviewed.